1 Fully Integrated Wearable Impedance Cytometry Platform on Flexible Circuit Board With Online Smartphone Readout
Bebe Schoenheimer edited this page 2025-08-30 00:57:03 +08:00
This file contains ambiguous Unicode characters

This file contains Unicode characters that might be confused with other characters. If you think that this is intentional, you can safely ignore this warning. Use the Escape button to reveal them.


The system diagram is displayed in Fig. 1. We use our customized-constructed analog architecture23, designed to detect extremely delicate impedance adjustments in a microfluidic channel with low-end hardware. Custom-built analog structure for impedance cytometry with off-the shelf hardware23. System block diagram of cytometer-readout structure. To carry out conventional LIA, a voltage at a high reference frequency is modulated with the microfluidic channel impedance, generating a current sign. The biosensor BloodVitals SPO2 used on this work relies on an electric field generated between two electrodes inside a microfluidic channel, with the baseline impedance representing phosphate buffered solution (PBS), and variable impedance resulting from particle flow via the electric discipline. A trans-impedance amplifier then amplifies the input present signal and outputs a voltage signal, which is then combined with the original reference voltage. Finally, a low-pass filter isolates the low-frequency part of the product, which is a low-noise sign proportional to the channel impedance amplitude on the reference frequency22.


As our channel impedance also varies with time, we designed the low-pass filter cutoff frequency to be bigger than the inverse of the transit time of the microfluidic particle, or the time it takes for the particle to transverse the sphere between electrodes. After performing conventional LIA on our biosensor, there stays a DC offset throughout the filtered signal which is along with our time-varying signal of curiosity. The DC offset limits the acquire that may be utilized to the sign before clipping happens, and in23, we describe the novel use of a DC-blocking stage to subtract the offset and apply a post-subtraction excessive-achieve amplification stage. The result is a extremely sensitive structure, which might be applied with a small footprint and off-the-shelf elements. For real-time SPO2 tracking an in-depth analysis on the structure, together with the noise evaluation and simulation, we refer to the unique work23. An vital observe is that the DC-blocking stage causes the positive voltage peak to be followed by a damaging voltage peak with the identical built-in energy, giving the novel structure a uniquely formed peak signature.


Because the analog signal has been amplified over several orders of magnitude, BloodVitals tracker a low-end ADC in a microcontroller chip can sample the info. The microcontroller interfaces with a Bluetooth module paired with a custom developed smartphone software. The application is used to initiate information sampling, and for information processing, readout and evaluation. We've carried out the architecture as a seamless and wearable microfluidic platform by designing a flexible circuit on a polyimide substrate in the type of a wristband (manufactured by FlexPCB, Santa Ana, CA, BloodVitals tracker USA) as shown in Fig. 2. All components, such as the batteries, microcontroller, Bluetooth module, BloodVitals home monitor and BloodVitals tracker biochip are unified onto one board. The versatile circuit is a two-layer polyimide board with copper traces totaling an area of eight in². Surface-mount-packaged elements were chosen to compact the general footprint and reduce noise. Lightweight coin cell lithium ion polymer (LIPO) batteries and regulator chips (LT1763 and LT1964 from Linear Technology) were used to supply ±5 V rails.


A 1 MHz AC crystal oscillator (SG-210 from EPSON), D flip-flop (74LS74D from Texas Instruments) for frequency division, and BloodVitals SPO2 device passive LC tank was used to generate the 500-kHz sine wave 2 Volt Peak-to-Peak (Vp-p) sign, which is excited through the biosensor. The glass wafer performing because the substrate for the biosensor was cut across the PDMS slab with a diamond scribe to reduce the dimensions and was attached to the board through micro-hook-tape and micro-loop-tape strips. The electrodes of the sensor interfaced with the board via jumping wires which have been first soldered to the circuits terminals after which bonded to the sensors terminals with conductive epoxy. Removal of the PDMS sensor includes de-soldering the leaping wires from the circuit board, separation of the micro-hook strip adhered to PDMS sensor from the underlying micro-loop strip adhered to the board, and vice versa for the addition of another sensor. A DC-blocking capacitor was added prior to the biosensor to prevent low-frequency energy surges from damaging the biosensor whereas the circuit was being switched on or off.


The trans-impedance stage following the biosensor was carried out with a low-noise operational amplifier (TL071CP from Texas Instruments) and a potentiometer within the suggestions path for adjustable achieve from 0.04 to 0.44. Mixing was achieved with a multiplier (AD835 from Analog Devices). To isolate the part of curiosity from the product of the mixing stage, a 3rd order Butterworth low-go filter with a one hundred Hz cutoff frequency and 60 dB roll off per decade was designed with another TL071CP op-amp23. A DC-blocking capacitor was used for BloodVitals SPO2 the DC-blocking stage. The last stage of the analog design, BloodVitals tracker the high acquire stage, was achieved with two extra TL071CP amplifiers. An ATtiny 85 8-bit microcontroller from Atmel driven by an external 16 MHz on-board crystal was used to pattern knowledge. The HM-10 Bluetooth Low Energy (BLE) module was used for knowledge transmission to the smartphone, with the module and BloodVitals tracker the breakout circuit built-in on-board. The process used to microfabricate our PDMS microfluidic channel for impedance cytometry is a regular one and has been beforehand BloodVitals tracker reported27.